冰研應用股份有限公司Advanced HiVac Application Co., LTD
  

V701 Turbo Pump



V701Manual.pdf

Pumping speed CF 10” or ISO 200: N2: 690 l/s He: 620 l/s H2: 510 l/s
Compression ratio N2: 1 x 109 He: 1 x 107 H2: 1 x 106
Base pressure* 
(with minimum recommended forepump)
<1 x 10-10 mbar (< 1 x 10-10 Torr)
Inlet flange CF 10” / ISO 200 / ISO 200-F bolted
Foreline flange KF25
Rotational speed 42,000 rpm
Start-up time < 5 minutes
Recommended forepump Mechanical: Varian DS 402 
Dry scroll: Varian TS300
Operating position Any
Cooling requirements Natural air convection
Water optional (use water with electrical
conductivity ≤ 500 μS/cm)
Bakeout temperature 120 °C at inlet flange max. (CF flange) 
80 °C at inlet flange max. (ISO flange)
Vibration level (displacement) <0.01 μm at inlet flange

<(lbs) kg>

ISO flange 19.4 (43.0) 
CF flange 25.5 (54.2) 
CF and ISO bolted flange 25.5 (54.2)


 

 

 

 

回上一頁