Technical Specifications
Pumping Speed, Operating Range Pumping Speed, AVS 4.1 (1963) | 4,500 l/s Air, 5,625 l/s He/H2 8,000 l/s Air |
Maximum Throughput | 6.3 T-l/s (8.4 mbar l/s) in operating range 7.5 T-l/s (10.0 mbar l/s) @ 0.01 torr |
Operating Range | 1.4 x 10-3 to <5 x 10-9 torr (1.9 x 10-3 to <6.5 x 10-9 mbar) |
Maximum Forepressure | No Load: 0.65 torr (0.85 mbar) Full Load: 0.55 torr (0.72 mbar) |
Recommended Backing Pump | ≥30 cfm (51 m3/hr) |
Backstreaming Rate, Standard Cold Cap | <1.0 x 10-3 mg/cm2/min |
Warmup Time | 15 minutes |
Cooldown Time | 25 minutes |
Fluid Charge | 1,000 cc |
Electrical Requirements | 3 ph, 50/60 Hz, 208/380/480 VAC |
Pump Power | 4400 watts |
Cooling Water Requirements | 0.40 gpm (80 l/hr) |
Water Connections |
|